Air to air temperature shock test chamber with fixed test space ShockEvent D1 / 3

Air to air temperature shock test chamber with fixed test space ShockEvent D

Air to air temperature shock test chamber with fixed test space ShockEvent D

ShockEvent D damper-type shock test chambers with fixed test space are ideal for vibration-sensitive or active test objects such as sensors and other electronic components.

ShockEvent D damper-type shock test chambers with fixed test space are ideal for vibration-sensitive or active test objects such as sensors and other electronic components. During the thermal cycling test, temperature-controlled hot or cold air flows into the test chamber to the test specimen. This means that the test specimen remains stationary and does not have to be moved between several temperature chambers, as is the case with horizontal and vertical shock test chambers.

This makes the damper-type shock test chamber unique.

  • Ideal for vibration-sensitive and active test specimens: The fixed test chamber allows for optimal stress conditions without having to move the test specimen. Cables and test setups can be easily fed through the side wall and remain stationary, just like the test specimen.
  • 3-Zone Temperature Shock: The thermal shock test method, which uses three temperature zones (cold, hot, ambient) within a single chamber, offers greater flexibility in testing procedures than conventional thermal shock chambers.
  • Efficient energy-saving mode: In energy-saving mode, the cold and hot chambers are temporarily turned off during programs with long dwell times. This significantly reduces energy consumption.
  • Comprehensive product safety: Multi-level safety mechanisms, such as automatic shutdown of the test specimen, over-temperature protection, and lockable doors, protect both operators and the test specimen.
  • Easy access to the test material: The standard 50 mm feed-through and the large test chamber door ensure easy installation of the test specimen as well as convenient loading and unloading.

How does the damper shock work?

In contrast to other shock test chambers, the ShockEvent D damper-type shock test chamber works with a fixed test space. Temperature-controlled hot and cold air is pre-tempered in chambers above and below the test chamber. The damper-style shock test chamber allows this air to flow into the test space via flaps. This makes it significantly different from horizontal and vertical shock test chambers, in which the test specimen moves between two or three temperature ranges in a test basket.

We develop the optimal solution for your application.

The ShockEvent D is designed for performing common thermal shock test standards.

Shock operation

EIAJ ED-2531B Na

IEC 60068-2-14 Na

JASO D 014-4

MIL-STD 202H, Meth. 107

MIL-STD 883L -2, Meth. 1010.9 (A, B)

By preconditioning the air in the hot and cold chambers, the temperature settling times can be reduced.

Options

  • Rectangular access port 160 x 50 mm to the left: The rectangular access port (160 x 50 mm) is installed in the left-hand wall. It allows the insertion of larger or several plugs and cables that do not fit through the standard access port.
  • Insertion basket: Additional wire mesh shelves can be inserted for the insertion of the test specimens (max. 7 shelves). Max. load per shelf 7.5 kg. A total load of 50 kg must not be exceeded.
  • Electric door tumbler, normally closed (NC): The test space door is unlocked when a test ends, is stopped, is paused or in the event of a malfunction in a test, within a temperature range of +10 °C to +40 °C, and can then be opened. The test space heaters, the test space fan and the potential-free contact are then switched off.
  • Temperature measurement on the test specimen: Many norms, such as the new version of IEC 60068-2-14 Nb, require close observation and recording of the temperature of the test specimen during the test phases. Our test chamber can be equipped with additional temperature sensors, the values from which are displayed and recorded on the display (via USB or with S!MPATI®). One of the sensors can also be used as a control sensor if required.
  • Analog measured value card 4 Pt100 inputs and 5 outputs: For the processing and output of analog measurement signals, five outputs 0 to 10 V and four inputs for Pt 100 are available.
  • Insulation of water inlet at a water flow of <+12 °C: Cooling water hoses in the test chamber are additionally insulated.
  • Hose set for cooling water network 1", 2x2.5m, flexible: For connection to a cooling water network, two flexible hoses with a G 1" connection and a length of 2.5 m are supplied.
  • Hose set for cooling water network 1", 2x5m, flexible: For connection to a cooling water network, two flexible hoses with a G 1" connection and a length of 5 m are supplied.

Available models

  • ShockEvent D/210/a/V1
  • ShockEvent D/210/e/V1