Gas probe for various gases
Ammonia. Nitrogen dioxide. Nitrogen oxide. Chlorine gas. Sulfur dioxide. Hydrogen sulfide. Ethylene oxide.
Technical data and functions
- Measurement of gas concentration in air, with multiple ranges and model variants.
- Available for ammonia (0 to 250 ppm), nitrogen dioxide (0 to 30 ppm), nitrogen oxide (0 to 50 ppm), chlorine gas (0 to 50 ppm), sulfur dioxide (0 to 250 ppm), hydrogen sulfide (0 to 250 ppm) and ethylene oxide (0 to 100 ppm).
- Measuring principle: electrochemical reaction, temperature compensated.
- Error of measured value ±3 % of full scale value, linearity < 2 % of full scale value, settling time t90 < 60 s.
- Transverse sensitivity < 2 % by integrated filter, zero point drift < 2 % per year.
- Operating conditions -10 to +40 °C and > 15 to 90 % r.h. non-condensing; life span of the measuring cell approx. 2 years typical.
- Output 4 to 20 mA and 24 V supply via the ALMEMO® connector, with 1.5 m connecting cable.
- Measuring head Ø 80 mm, height 80 mm, weight 600 g; operation with the device in sleep mode is not possible.