Humidity Test Cabinets Alpha+
Alpha+ H500 and H1000 humidity cabinets for continuous condensation humidity tests at a single temperature and temperature cycling tests, such as DIN 50017, ISO 6270-1 and ASTM D2247.
H500 & H1000 Humidity cabinets meet the requirements of basic, continuous condensation humidity tests conducted at a single temperature and temperature cycling tests such as DIN50017, ISO 6270-1, VDA 621-421, ASTM D4585, ASTM D2247 and BS3900 and similar international test standards.
Alpha+ Condensation-Based Climate Chamber with Passive Moisture Saturation.
Precise temperature controlled, high humidity environment. Water held in a sump located in the cabinet base, is heated by an immersion heater to generate water vapour. As water vapour evaporates, it saturates the cabinet atmosphere with moisture. Test samples in the cabinet, below the dew-point temperature of the saturated air, will develop surface condensation. By careful design the high humidity environment is achieved without the need for a circulation fan, which would otherwise have to endure the hostile, high humidity environment.
Alpha+ - Humidity Test Chambers Standard.
User can program either a single set temperature, or automatically cycle the cabinet temperature between two different temperatures for user defined periods of time.
Standard models are designed for continuous humidity tests, such as DIN 50 017-KK, DIN 50 017-KFW, DIN 50 017-KTW, ISO 6270-2, VDA 621-421, ASTM D2247, BS 3900 Part F2.
Installation requirements.
Water: Deionized/distilled for topping up condensate humidity sump.
Air: Clean dry & oil free, 4.0 to 6.0 bar (58-87psi) with 240 Ltrs (8.5cu.ft) per minute flow
Exhaust: 3m (10ft) exhaust pipe is required which should be terminated outside building (see accessory ACC58).
Drain: 3m (10ft) drain pipe is required which should be terminated into floor level drain (see accessory ACC58).
Operating environment conditions: +18 to +23°C (+64 to 73°F), 85% max RH (non condensing) ambient
Options
- Cleveland Roof for Humidity Cabinet (ACC48): With the Cleveland style roof the test panels provide the actual roof of the cabinet, room air on the top side of the panel cools them to a couple of degrees below the vapour temperature, this temperature difference allows the condensate to form directly on the underside of the test panels.
- Access Port 35/110mm (ACC10): A 35mm (1.38") or 110mm (4.33") diameter sealable chamber access port. Passes through the left-hand wall of the chamber to enable the connection of external monitoring/driving equipment.
- Waste Water Trough & Pump System (ACC20): This option comprises of a custom made holding tank, which locates on the floor immediately behind the chamber, into which condensate from the chamber drain can be directed. When the holding tank fills to a pre-determined level, the condensate will automatically be pumped out by a low voltage marine specification pump, with power being supplied from the Ascott chamber via a panel mount socket and plug. This will discharge waste water to a remote drain or effluent tank located up to 10M/30ft away horizontally and up to 3M/10ft vertically.
Available models
- H500
- H1000
La fel la toate modelele
- Cabinet temperature:
- ambient +10 °C … +55 °C
- Power supply:
- 220 … 240 V, 50/60 Hz, 1 phase, max. 10 A
Testspace volume2 valori
- 500 l
- H500
- 1000 l
- H1000
Test Space Dimensions (H x W x D)2 valori
- 866 × 1076 × 683 mm
- H500
- 970 × 1553 × 863 mm
- H1000
External Dimensions (H x W x D)2 valori
- 1102 × 1650 × 934 mm
- H500
- 1159 × 2127 × 1127 mm
- H1000
Total Weight2 valori
- 130 kg
- H500
- 190 kg
- H1000
Load capacity2 valori
- 80 kg
- H500
- 150 kg
- H1000
Sample holders2 valori
- 5, up to 150 test panels
- H500
- 7, up to 266 test panels
- H1000
| Specificație | H500 | H1000 |
|---|---|---|
| Testspace volume | 500 l | 1000 l |
| Test Space Dimensions (H x W x D) | 866 × 1076 × 683 mm | 970 × 1553 × 863 mm |
| External Dimensions (H x W x D) | 1102 × 1650 × 934 mm | 1159 × 2127 × 1127 mm |
| Total Weight | 130 kg | 190 kg |
| Load capacity | 80 kg | 150 kg |
| Sample holders | 5, up to 150 test panels | 7, up to 266 test panels |
| Cabinet temperature | ambient +10 °C … +55 °C | |
| Power supply | 220 … 240 V, 50/60 Hz, 1 phase, max. 10 A | |