PharmaEvent
Medicinal products are subject to particularly strict guidelines in order to ensure patient safety and efficacy.
Consistent compliance with stability tests
Medicinal products are subject to particularly strict guidelines in order to ensure patient safety and efficacy. With our stability testing systems, you can determine the conditions under which pharmaceutical substances have the longest shelf life in accordance with compliance regulations. The spectrum ranges from laboratory-scale systems to walk-in climate chambers for optimum long-term testing. The functionality of all systems is based on a risk-based approach and with all the necessary device functions for ideal GMP compliance.
Overview
The PharmaEvent series was specially developed to meet the requirements of stability testing laboratories in the pharmaceutical industry. The systems are available in four sizes and ensure a constant climate (types C/280, C/600, C1300 and C/2000) or a constant temperature (types T/280, T/600, T/1300 and T/2000). They can also be used for changing climates. The robust design, innovative product functions, precise control system, and intelligent controls enable stability testing with maximum safety and ease of use. The work area easily meets the requirements of ICH guideline Q1A. Documentation according to FDA 21 CFR Part 11 and EU GMP Annex 11 and compliance with ICH guidelines Q1A and Q1B as well as the requirements of WHO, FDA, and CPMP are a matter of course for us.
Stability testing, easier and safer than ever
- Absolute homogeneity thanks to ideal air flow – even with loaded devices
- Sterile steam system for killing any microorganisms that may be present
- GMP-compliant user management with AuditTrail
- Alarm system according to GAMP
- Operation from the workstation without additional software using the integrated WEBSeason®
- Multilingual operation
- Connection to our Simpati Pharma monitoring system; can also be integrated into LIMS systems
Functionality
The stability test chambers in the PharmaEvent series are designed so that continuous operation at 5°C is possible without defrosting. They also allow tests to be carried out with other specifications in the performance range of the respective system. The temperature and humidity are regulated using high-precision sensors in conjunction with a specially developed control system. The control system reacts quickly to correct set-point deviations caused by:
- Influence of the chamber contents (absorption or release of water vapor by the test material or its packaging)
- External influences (e.g., laboratory temperature, door opening)
Best Practice
The highly accurate and reliable PharmaEvent stability test chambers are validatable and specifically designed to enable you to meet the requirements of ICH guideline Q1A. Accuracy is also ensured taking into account the physical deviation and the measurement uncertainty of the measuring and monitoring systems.
La fel la toate modelele
- Insert grille, W × D:
- 530 × 650 mm
- Insert grille, Load per slide-in unit:
- 40 kg
- Temperature, Working range:
- +2 to +70 °C
- Temperature, Deviation:
- ±0.1 to ±0.2 K
- Temperature, Homogeneity, spatial:
- ±0.3 to ±1.0 K
- Temperature, Gradient according to IEC 60068-3-5:
- 0.5 to 2 K
- Humidity [% r.h.]:
- 20 to 90
- Humidity, Deviation [% r.h.]:
- ±0.2 to ±1.0
- Humidity, Dew point temperature range:
- +5 to +45 °C
- Calibration values (factory calibration):
- +25°C/60% r.h. and 40°C/75% r.h.
Insert grille, Number (max.)4 valori
- 2 (17)
- C/280
- 6 (36)
- C/600
- 12 (77)
- C/1300
- 18 (108)
- C/2000
Insert grille, Storage area (max.) [m²]4 valori
- 0.69 (5.86)
- C/280
- 2.07 (12.4)
- C/600
- 4.14 (26.5)
- C/1300
- 6.21 (37.2)
- C/2000
Insert grille, Total load4 valori
- 150 kg
- C/280
- 250 kg
- C/600
- 500 kg
- C/1300
- 750 kg
- C/2000
Outside dimensions, W × D × H4 valori
- 1159 × 872 × 1017 mm
- C/280
- 803 × 1060 × 1995 mm
- C/600
- 1523 × 1043 × 1995 mm
- C/1300
- 2180 × 1040 × 2000 mm
- C/2000
Test space dimensions, W × D × H4 valori
- 645 × 673 × 641 mm
- C/280
- 621 × 687 × 1280 mm
- C/600
- 1341 × 687 × 1280 mm
- C/1300
- 2035 × 695 × 1280 mm
- C/2000
Installed power2 valori
- 1.1 kW
- C/280
- 1.2 kW
- C/600, C/1300, C/2000
| Specificație | C/280 | C/600 | C/1300 | C/2000 |
|---|---|---|---|---|
| Insert grille, Number (max.) | 2 (17) | 6 (36) | 12 (77) | 18 (108) |
| Insert grille, Storage area (max.) [m²] | 0.69 (5.86) | 2.07 (12.4) | 4.14 (26.5) | 6.21 (37.2) |
| Insert grille, Total load | 150 kg | 250 kg | 500 kg | 750 kg |
| Outside dimensions, W × D × H | 1159 × 872 × 1017 mm | 803 × 1060 × 1995 mm | 1523 × 1043 × 1995 mm | 2180 × 1040 × 2000 mm |
| Test space dimensions, W × D × H | 645 × 673 × 641 mm | 621 × 687 × 1280 mm | 1341 × 687 × 1280 mm | 2035 × 695 × 1280 mm |
| Installed power | 1.1 kW | 1.2 kW | ||
| Insert grille, W × D | 530 × 650 mm | |||
| Insert grille, Load per slide-in unit | 40 kg | |||
| Temperature, Working range | +2 to +70 °C | |||
| Temperature, Deviation | ±0.1 to ±0.2 K | |||
| Temperature, Homogeneity, spatial | ±0.3 to ±1.0 K | |||
| Temperature, Gradient according to IEC 60068-3-5 | 0.5 to 2 K | |||
| Humidity [% r.h.] | 20 to 90 | |||
| Humidity, Deviation [% r.h.] | ±0.2 to ±1.0 | |||
| Humidity, Dew point temperature range | +5 to +45 °C | |||
| Calibration values (factory calibration) | +25°C/60% r.h. and 40°C/75% r.h. | |||
Valoare comună modelelor peste care se întinde