PharmaEvent walk-in stability test chambers
Medicinal products are subject to particularly strict guidelines worldwide.
Consistently uncompromising in stability tests in almost unlimited chamber sizes
Medicinal products are subject to particularly strict guidelines worldwide. With our stability testing systems, you can determine the shelf life of pharmaceutical substances extremely safely and in accordance with legal requirements. The spectrum ranges from laboratory-scale systems to walk-in climate chambers for optimum long-term testing. All systems have the necessary documentation options according to FDA 21 CFR Part 11 and EU GMP Annex 11 and comply with ICH guidelines Q1A and Q1B.
Overview
The highly accurate and reliable PharmaEvent stability test chambers are validatable and specifically designed to enable you to meet the requirements of ICH guideline Q1A. Also other national and international requirements such as WHO, FDA, and CPMP are met. The insulating cells can be optimally adapted to the existing building structure because there are no fixed dimensions. The standard height is 2,700 mm; other dimensions are possible. Test chambers and test spaces with a volume of 10–800 m³ are available.
Stability testing, easier and safer than ever
- Ideal for integration into your building
- Absolute homogeneity thanks to ideal air flow – even with loaded systems
- Energy-efficient construction
- GMP-compliant user management with audit trail and alarm system according to GAMP
- Operation from the workstation without additional software using the integrated WEBSeason®
- Multilingual operation
- Connection to the Simpati pharma monitoring system; can also be integrated into LIMS systems
Functionality
- Excellent mechanical strength and optimum thermal insulation thanks to PU insulation cells (CFC-free) with easy-to-clean metal plate coating on both sides. Depending on the application, the walls are made of the appropriate material (e.g., aluminum, stainless steel).
- Insulated base construction for heavy-duty use with non-slip stainless steel checker plate.
- Lockable test space door with isolated observation window and emergency opening device. The door frame heater prevents condensate formation during operation with high humidity.
- Pressure compensation valve integrated into the cell wall.
- Temperature control system consisting of ceiling evaporator with integrated electric heater and air or water-cooled chiller unit.
- Powerful axial fans ensure the best possible air circulation as well as uniform air movement and tempering.
- Humidification system with energy-saving ultrasonic humidifier and separate dehumidifying evaporator.
- Fully integrated, closed simulation system for the best accuracies instead of an attached conditioning unit.
- Temperature range
- +20 to +45 °C
- Temperature deviation, Temporal
- ±0.1 to ±0.5 K
- Temperature homogeneity, spatial
- ±0.5 to ±1.0 K
- Temperature gradient, according to IEC 60068-3-5
- 1 to 2 K
- Humidity range [% r.H.]
- 20 to 80
- Humidity deviation, temporal [% r.h.]
- ±1 to ±3
- Dew point temperature range
- +9 to +41 °C