Photo stability test chambers
The photo-stability test chambers in the PharmaEvent series are available in two sizes. They generate either a constant climate (Type C) or only a constant temperature (Type T).
Ideal light distribution for the best reproducibility
The photo-stability test chambers in the PharmaEvent series are available in two sizes. They generate either a constant climate (Type C) or only a constant temperature (Type T). The photo-stability test chambers are characterized by an ideal light, UV, temperature, and humidity distribution (type C) and thus guarantee absolutely reproducible light, UV, and climate conditions. The lighting equipment used meets the requirements of ICH guideline Q1B Option 2 and enables photo-stability tests to be carried out in less than 100 h.
Overview
One of the most important requirements for photo-stability tests is the homogeneous irradiation of the samples. It is therefore essential to position all samples at the same distance from the light source. The inhomogeneous light emission of fluorescent tubes is balanced out by special light and UV filter systems, thereby resulting in homogeneous illumination of the entire storage area. This system can be equipped with appropriate light and UV sensors to record the illuminance and UV irradiance. This option allows the set-points to be entered in lxh and Wh/m² (e.g., 1.2 million lxh and 200 Wh/m²), thereby providing a fully automated and, in conjunction with S!MPATI® pharma, fully documented process. PharmaEvent photo-stability test chambers offer innovative product functions, high accuracy, intelligent control systems, and high-quality construction.
The best reproducibility; photo-stability tests are easier and safer than ever
- The best reproducibility thanks to ideal light distribution with patented light filters
- Fully automated process thanks to light integration with optional sensors
- Sterile steam system for killing any microorganisms that may be present
- GMP-compliant user management with AuditTrail
- Operation from the workstation without additional software using the integrated WEBSeason®
- Multilingual operation
- Connection to our Simpati Pharma monitoring system; can also be integrated into LIMS systems
Functionality
The highly accurate and reliable PharmaEvent photo-stability test chambers are validatable and specifically designed to enable you to meet the requirements of ICH guideline Q1B. For this purpose, proof of sufficient light homogeneity is essential.
Accuracy is also ensured taking into account the physical deviation and the measurement uncertainty of the measuring and monitoring systems.
La fel la toate modelele
- Insert grille, Load per slide-in unit:
- 25 (distributed load) kg
- Temperature, Working range:
- without irradiation: +5 to +50 °C; with irradiation: +5 to +50 °C (250 l) / +10 to +50 °C (500 l)
- Temperature, Deviation:
- ±0.1 to ±0.5 K
- Temperature, Homogeneity, spatial:
- ±0.5 to ±1.0 K (1.5 K with radiation)
- Temperature, Gradient according to IEC 60068-3-5:
- 1 to 2 K
- Light, Illuminance [lx]:
- approx. 5,500 at 5°C (only 250 l); approx. 12,000 at 15°C; approx. 18,000 at 25°C; approx. 25,000 at 45°C
- Light, UV illuminance [W/m²]:
- 0.65 at 5°C (only 250 l); 1.5 at 15°C; 3.0 at 25°C; 3.7 at 45°C
- Light, Homogeneity:
- ±0.5 to ±1.0 K (1.5 K with radiation)
- Light, Light distribution:
- approx. ±8 %
- Light, UV distribution:
- approx. ±12 %
- Calibration values (factory calibration):
- +25°C/60% r.h. and 40°C/75% r.h.
Insert grille, Number (max.)2 valori
- 2: 1 UV; 1 cold white light
- C/250/L, T/250/L
- 4: 2 UV; 2 cold white lights
- C/500/L, T/500/L
Insert grille, Storage area [m²]2 valori
- 0.71
- C/250/L, T/250/L
- 1.45
- C/500/L, T/500/L
Insert grille, Total load2 valori
- 50 kg
- C/250/L, T/250/L
- 100 kg
- C/500/L, T/500/L
Outside dimensions, W × D × H2 valori
- 1159 × 872 × 1017 mm
- C/250/L, T/250/L
- 803 × 1060 × 2055 mm
- C/500/L, T/500/L
Test space dimensions, W × D × H2 valori
- 530 × 673 × 641 mm
- C/250/L, T/250/L
- 530 × 687 × 1305 mm
- C/500/L, T/500/L
Test space dimensions, Usable test space, approx.2 valori
- 235 l
- C/250/L, T/250/L
- 460 l
- C/500/L, T/500/L
Weight2 valori
- 161 kg
- C/250/L, T/250/L
- 250 kg
- C/500/L, T/500/L
Humidity [% r.h.]2 valori
- 20 to 90
- C/250/L, C/500/L
- —
- T/250/L, T/500/L
Humidity, Deviation in time [% r.h.]2 valori
- ±1 to ±2
- C/250/L, C/500/L
- —
- T/250/L, T/500/L
Humidity, Dew point temperature range2 valori
- +5 to +40 °C
- C/250/L, C/500/L
- —
- T/250/L, T/500/L
Installed power2 valori
- 1.4 kW
- C/250/L, T/250/L
- 2.6 kW
- C/500/L, T/500/L
| Specificație | C/250/L | C/500/L | T/250/L | T/500/L |
|---|---|---|---|---|
| Insert grille, Number (max.) | 2: 1 UV; 1 cold white light | 4: 2 UV; 2 cold white lights | 2: 1 UV; 1 cold white light | 4: 2 UV; 2 cold white lights |
| Insert grille, Storage area [m²] | 0.71 | 1.45 | 0.71 | 1.45 |
| Insert grille, Total load | 50 kg | 100 kg | 50 kg | 100 kg |
| Outside dimensions, W × D × H | 1159 × 872 × 1017 mm | 803 × 1060 × 2055 mm | 1159 × 872 × 1017 mm | 803 × 1060 × 2055 mm |
| Test space dimensions, W × D × H | 530 × 673 × 641 mm | 530 × 687 × 1305 mm | 530 × 673 × 641 mm | 530 × 687 × 1305 mm |
| Test space dimensions, Usable test space, approx. | 235 l | 460 l | 235 l | 460 l |
| Weight | 161 kg | 250 kg | 161 kg | 250 kg |
| Humidity [% r.h.] | 20 to 90 | — | — | |
| Humidity, Deviation in time [% r.h.] | ±1 to ±2 | — | — | |
| Humidity, Dew point temperature range | +5 to +40 °C | — | — | |
| Installed power | 1.4 kW | 2.6 kW | 1.4 kW | 2.6 kW |
| Insert grille, Load per slide-in unit | 25 (distributed load) kg | |||
| Temperature, Working range | without irradiation: +5 to +50 °C; with irradiation: +5 to +50 °C (250 l) / +10 to +50 °C (500 l) | |||
| Temperature, Deviation | ±0.1 to ±0.5 K | |||
| Temperature, Homogeneity, spatial | ±0.5 to ±1.0 K (1.5 K with radiation) | |||
| Temperature, Gradient according to IEC 60068-3-5 | 1 to 2 K | |||
| Light, Illuminance [lx] | approx. 5,500 at 5°C (only 250 l); approx. 12,000 at 15°C; approx. 18,000 at 25°C; approx. 25,000 at 45°C | |||
| Light, UV illuminance [W/m²] | 0.65 at 5°C (only 250 l); 1.5 at 15°C; 3.0 at 25°C; 3.7 at 45°C | |||
| Light, Homogeneity | ±0.5 to ±1.0 K (1.5 K with radiation) | |||
| Light, Light distribution | approx. ±8 % | |||
| Light, UV distribution | approx. ±12 % | |||
| Calibration values (factory calibration) | +25°C/60% r.h. and 40°C/75% r.h. | |||