Reach-In Temperature & Climate Test Chamber Temp-& ClimeEvent
Our ClimeEvent climate chamber is as innovative as it is sophisticated and offers you real added value in everyday testing.
Our ClimeEvent climate chamber is as innovative as it is sophisticated and offers you real added value in everyday testing. With increased humidification and dehumidification performance and a new type of control for even more precise humidity values - the perfect solution for testing the influence of temperature and humidity on the properties, function and service life of your products.
This makes the ClimeEvent unique.
- Also with new climate-friendly refrigerant R744: Our single-stage systems (down to -40°C) are equipped with CO2 refrigeration technology (GWP = 1). Even at -40°C, the system still offers sufficient power to operate your test material.
- Connectivity is key: Versatile thanks to our in-house Simpati software. Flexibility thanks to various interfaces (SimServ, interface, LabVIEW, OPC-UA, ProfiBus, ProfiNET or EtherCAT). ASCII can still be used.
- Even more precise humidity values and simple management of the humidification water: Innovative absolute humidity control ensures control accuracy for climate change tests and extreme temperature and humidity combinations. Large water bath for increased humidification and dehumidification performance. 25-litre water tank - automatic make-up included as standard.
- Protect your valuable test material: Triple test material fuse included as standard. Active test items can be deenergized via a potential-free contact. Test chamber can switch off in the event of a fault in the test material or be set to a defined value (optional).
- Requirements from the defense sector: Whether you need to meet MIL standards or implement special applications such as the temperature control of ammunition in external test housings - our products offer solutions for a wide range of applications.
- End-of-line testing: For 100% testing after production, particularly efficient loading and unloading are relevant, as well as integration into existing systems with OPC-UA or SECS/GEM. We offer individual solutions for this.
The ClimeEvent now also with climate-friendly refrigerant CO2 (R744).
Our single-stage systems (down to -40°C) are equipped with the innovative CO2 refrigeration technology (GWP = 1).
- GWP = 1 - future-proof testing!
- Our CO2 systems provide powerful cooling down to -50°C.
- No changes to the infrastructure necessary.
- Compared to cascade units (down to -70°C), often lower investment and operating costs.
- Lower energy consumption, with ramp programmes between 120 °C and -40 °C.
- For example, a CO2 test chamber consumes up to 30 % less energy than a cascade.
- Significantly reduced sound pressure level.
We develop the optimum solution for your application.
Automotive: Demanding climate tests such as those from LV124 or general standards such as the IEC 60068 series pose no challenges for our test chambers. Not even in the field of e-mobility.
Electronics: Our test chambers are designed for a wide range of tests in the electronics industry and easily master these specific requirements; e.g. low-silicone test chamber.
5G and autonomous driving: There are often very specific requirements in this area. Special adaptations enable testing in the EMC area or of lidar sensors.
Aerospace and defence: In this area, extreme demands are often placed on the product. Our chambers can be used to test corresponding standards from the aerospace or defence sector.
Consumer products: Consumer and luxury goods must also be tested. These sometimes pose special challenges for the test chambers, such as tests with leather, which introduces highly corrosive outgassing into the test chamber. Here too, we have already provided solutions for our customers on several occasions.
Special applications.
Lithium ions: Rechargeable batteries are installed in many devices these days, and e-mobility is also leading to an increased number of tests with batteries. To ensure that you can test these test items safely, our test chambers are equipped with a tested safety system based on the EUCAR Hazard Level System, which extends up to Hazard Level 7.
ATEX: Many test items use flammable substances as lubricants or operating fluids. A risk assessment is required in order to be able to test safely even in the event of a leak in the test material. If necessary, measures must be taken at the test chamber. We offer you advisory support and equip your test chamber with a tested safety concept.
ESD: Some test items are very sensitive to electrostatic charge. The test chambers can be equipped accordingly to prevent this from happening in the test chamber, for example through air movement. From the earthing point to the shielding of all installed components.
EMC: The test chambers must be specially shielded to ensure that no signals are emitted from the test object into the environment or affect your test object. We have many years of experience in the construction of test chambers for EMC tests.
Corrosive gases: PCBs and plug connections in particular must be tested under conditions with corrosive gases. Appropriate equipment is also available for this purpose.
Stress screening test chambers for temperature and climatic tests
The Reach-In temperature and climate test chambers for stress screening tests are available as standard in the performance classes 10 K/min to 25 K/min. The ideal test chamber for the test specimen can be selected from sizes ranging from 270 to 1300 liters. Our ESS test chambers are specially developed for fast temperature cycles and moist heat tests and are used primarily in safety-critical sectors such as the aerospace, automotive, medical technology and defense industries.
Options
- Additional Access Ports: Access port made of stainless steel, available in three standard sizes Ø 50, 80, 125 mm, including 1 silicone plug closed and 1 foam + silicone plug, slotted.
- Heavy-load rails: This option includes a modification of the test space and the exterior housing. Via special rails, the test specimen load is transferred from the test space to the exterior housing. Depending on the size, the surface load is up to 350 kg or 500 kg. The test space height is reduced by 5°mm.
- Web panel below the door lock: The operating/programming and monitoring unit with a 25.4 cm (10") web panel is mounted below the door lock.
- Webpanel installed at any location on side wall: The operating/programming and monitoring unit with 25.4cm (10") web panel can be mounted anywhere on the side panel. (The exact positioning must be specified when ordering)
- Demineralization unit, 10 l, pressure-proof to 6 bar: If no demineralized water is available from the customer, a demineralization unit can be connected upstream. ( Replacement cartridge available)
- Activated carbon filter for demineralized water: Fitting located behind the lifting gear. The standard access port (Ø 80 mm) remains installed.
- Access port in the ceiling: Access port in the ceiling made of stainless steel, available in three standard sizes Ø 50, 80, 125 mm.
- Lead-through pad, incl. replacement insert: The installation is located behind the lifting gear. The standard feedthrough, Ø 80 mm, stays installed.
- Notched access port welded: A welded notch bushing measuring approx. 50 mm x 50 mm is fitted in the housing cover on the right for inserting cables. This allows test material to be inserted without having to remove cables and lines.
- Disconnection of fan via digital channel: Disconnection of the fan is activated and deactivated via a digital channel on the control panel. If the digital channel “disconnection of fan” is activated, the fan and the temperature-conditioning unit are switched off immediately.
- Window in the door: Multi-insulated, heated viewing window in the door. (Format: 520 mm x 418 mm)
- Switchable privacy screen for test specimens: You want to protect your test specimen comfortably from prying eyes? The window pane of the test chamber can be made opaque by means of a digital switching channel via the light switch. (Only in connection with the “window in the door” option)
- Door hinged on the right: Version of the device door with door hinge on the right side. The notched port option is not possible. The flat notch port option is possible at the top left.
- Door seals can be replaced for tests with hydraulic oil: Hydraulic oil places special demands on the seals. They have to be replaced more often. The seals are mounted on the door to make them easier to replace.
- Test chamber activation via digital input: The test can only be started if a voltage signal of +24 V DC max. 30 mA is present at the digital input or if the adapter plug has been plugged into the 25-pin D-Sub socket “Digital I/O”.
- Emergency-stop button on test space housing: The emergency stop button is installed on the outside of the test chamber, at the height of the door lock. When actuated, a message is displayed on the control panel and the test is stopped.
- Signal lamp, two colours: The two-colour signal lamp installed on top of the test chamber indicates the operating mode.
- Fault message to potential-free switching contact: If a fault occurs at the test chamber, a potential-free switching contact is actuated.
- Door contact switch for “Door open” notification on the control unit / S!MPATI®: The components of the door contact switch are mounted on the test chamber and on the test chamber door. Monitoring of the test space door is activated when the test starts. When the test space door is opened, the message “Door open” appears on the control panel.
- Safety interlock switch, open at zero current: The components of the electrical door interlock are mounted on the test chamber and on the test chamber door. The monitoring of the test chamber door is activated when the test is started.
- Safety interlock switch, closed at zero current: The components of the electrical door lock are mounted on the test chamber and test chamber door. The control system for the test chamber door will be activated when starting the test.
- Extension of climate range with compressed air: With the optional “capacitive humidity measuring system” (mat. no. 64624928) controlled operation with a dew point down to -20 °C is possible if compressed air is provided by the customer.
- Unregulated compressed air dryer for dew points down to -30 °C: Dehumidification equipment for dew points down to -30 °C (unregulated operation) to prevent condensation on the test specimen during temperature tests. Activation and deactivation are controlled via a digital switching channel. For climate test chambers, regulated operation is possible up to a dew point of -12 °C.
- Compressed air dryer for dew points down to -20 °C controlled / down to -30 °C uncontrolled: Dehumidification device for Climate Test Chambers including a capacitive humidity measuring system for dew points down to -20 °C (controlled operation). Controlled dehumidification is started by activating the dehumidification device via a digital switch channel. It can also be used as a dehumidification device for dew points down to -30 °C (uncontrolled operation) to avoid condensation formation on the test specimen during temperature tests.
- GN2 / Compressed air connection: Connection with coupling plug DN 7.2 for operation with the customer’s compressed air dryer or for the injection of an inert gas into the hot and cold chamber. The function can be enabled/disabled on the control unit or via the S!MPATI® software.
- Additional humidity control with capacitive humidity measuring system: In addition to the standard-issue psychrometric measuring device, a capacitive humidity measuring system is also installed. Humidity control is enabled either via this humidity sensor or the psychrometric measuring device.
- Independent humidity measuring sensor for reference measurement: Capacitive sensor for humidity measurement and sensor for temperature measurement, located in the supply air next to the control sensor. This option is used for reference measurement, as the measured values are recorded independently of the unit control.
- Temperature measurement on the test specimen: Many norms, such as the new version of IEC 60068-2-14 Nb, require close observation and recording of the temperature of the test specimen during the test phases. Our test chamber can be equipped with additional temperature sensors, the values from which are displayed and recorded on the display (via USB or with S!MPATI®). One of the sensors can also be used as a control sensor if required.
- Temperature measurement on test specimen (max. 1 sensor) switchable as control sensor: The changeover is via a digital switch channel. The measured value is displayed on the control unit and can be called up via the interfaces. The sensor is inserted into the test space via the standard port.
- Temperature measurement on test specimen: Movable temperature sensor Pt 100 with flexible cable for temperature measurement at any point in the test room or on the test object. The measured value is displayed in °C on the control panel and can be called up via the interfaces or the free analogue outputs of the measured value card.
- Temperature measurement on the test specimen using the NiCrNi sheath thermocouple: Movable temperature sensor as Ni-Cr-Ni thermocouple with flexible cable for temperature measurement anywhere in the test space or on the test specimen. The flexible cable length is approx. 5m and the sensor is 3m in length.
- Temperature measurement on the test specimen via NiCrNi thermocouple wire: Movable temperature sensor as Ni-Cr-Ni thermocouple wire, welded measuring tips, with a flexible cable for temperature measurement anywhere in the test space or on the test specimen. The flexible cable length is approx. 5 m.
- Temperature measurement at the test specimen by means of thermocouple Ni-Cr-Ni, switchable as control sensor: The switchover is via digital switch channel. The measured value is displayed on the control unit and can be called up via the interfaces.
- Temperature measurement on the test specimen via transmitter using a NiCrNi thermocouple: Movable temperature sensor as Ni-Cr-Ni thermocouple with flexible cable for temperature measurement anywhere in the test space or on the test specimen.
- Additional 4 Digital I/O: 4 digital outputs for controlling the additional functions via the potential-free contacts, load max. 24 V-DC, 0.5 A, 4 digital inputs can be queried via interface, load max. 24 V-DC, approx. 30 mA
- Analogue measuring card: 5 outputs 0 to 10 V and 4 inputs for Pt 100 are available for processing and output of analog measuring signals.
- Analogue measured value card: 5 outputs 0 to 10 V and 4 inputs for Ni-Cr-Ni thermocouple type K are available for the processing and output of analogue measurement signals.
- Energy meter: Professional energy analysis with an energy meter
- Air- or water-cooled: Many test chambers are available as air- or water-cooled and can be supplied according to your infrastructure. Water cooling is available for all systems and, for more powerful test chambers, it often makes more sense to dissipate the waste heat via a cooling water connection. Air-cooled versions are to be discussed on an individual basis on request.
- Supply connections installed at the top: The following supply connections are mounted on the top: Cooling water connection including optional insulation of water supply. Supply of demineralized water (Climate Test Chambers only) Compressed air supply (available as additional equipment)
- Hose kit for cooling water network 3/4", 2x2.5m, flexible: For connection to a cooling water network, two flexible hoses with a connection of G ¾” and a length of 2.5 m are provided.
- Hose kit for cooling water network 3/4", 2x5m, flexible: For connection to a cooling water network, two flexible hoses with a connection of G ¾” and a length of 5 m are provided.
- Hose set for cooling water network 1", 2x2.5m, flexible: For connection to a cooling water network, two flexible hoses with a G 1" connection and a length of 2.5 m are supplied.
- Hose set for cooling water network 1", 2x5m, flexible: For connection to a cooling water network, two flexible hoses with a G 1" connection and a length of 5 m are supplied.
- Insulation of water inlet at a water flow of <+12 °C: Cooling water hoses in the test chamber are additionally insulated.
- Pump system for floors without drain: The integrated pump-down system pumps the water in the system (condensate, humidification water, cleaning water) against the gradient into a drain on the customer side
- Special voltage on request: Special voltage on request
- Water Cooling CO2
Available models
- C2/180/50/3 (CO2)
- C2/180/70/3
- C2/180/70/5
- C2/340/50/3 (CO2)
- C2/340/70/3
- C2/340/70/5
- C2/600/50/4 (CO2)
- C2/600/70/3
- C2/600/70/5
- C2/1000/70/3
- C2/1000/70/5
- C2/1000/50/10 (CO2)
- C2/1500/70/3
- C2/1500/70/5
- C2/1500/50/7 (CO2)
- C2/2000/70/5
La fel la toate modelele
- Humidity range:
- 10 … 98 % R.H.
Testspace volume7 valori
- 190 l
- C2/180/50/3 CO2, C2/180/70/3, C2/180/70/5
- 335 l
- C2/340/50/3 CO2, C2/340/70/3
- 332 l
- C2/340/70/5
- 600 l
- C2/600/50/4 CO2, C2/600/70/3, C2/600/70/5
- 990 l
- C2/1000/70/3, C2/1000/70/5, C2/1000/50/10 CO2
- 1540 l
- C2/1500/70/3, C2/1500/70/5, C2/1500/50/7 CO2
- 2250 l
- C2/2000/70/5
External Dimensions (HxWxD)6 valori
- 1830 × 900 × 1575 mm
- C2/180/50/3 CO2, C2/180/70/3, C2/180/70/5
- 1830 × 900 × 1890 mm
- C2/340/50/3 CO2, C2/340/70/3, C2/340/70/5
- 2040 × 1120 × 1925 mm
- C2/600/50/4 CO2, C2/600/70/3, C2/600/70/5
- 2040 × 1420 × 2075 mm
- C2/1000/70/3, C2/1000/70/5, C2/1000/50/10 CO2
- 2040 × 1420 × 2600 mm
- C2/1500/70/3, C2/1500/70/5, C2/1500/50/7 CO2
- 2040 × 1420 × 3275 mm
- C2/2000/70/5
Test Space Dimensions (H x W x D)6 valori
- 750 × 580 × 450 mm
- C2/180/50/3 CO2, C2/180/70/3, C2/180/70/5
- 750 × 580 × 765 mm
- C2/340/50/3 CO2, C2/340/70/3, C2/340/70/5
- 950 × 800 × 800 mm
- C2/600/50/4 CO2, C2/600/70/3, C2/600/70/5
- 950 × 1100 × 950 mm
- C2/1000/70/3, C2/1000/70/5, C2/1000/50/10 CO2
- 950 × 1100 × 1475 mm
- C2/1500/70/3, C2/1500/70/5, C2/1500/50/7 CO2
- 950 × 1100 × 2150 mm
- C2/2000/70/5
Temperature range2 valori
- -50 … 180 °C
- C2/180/50/3 CO2, C2/340/50/3 CO2, C2/600/50/4 CO2, C2/1000/50/10 CO2, C2/1500/50/7 CO2
- -70 … 180 °C
- C2/180/70/3, C2/180/70/5, C2/340/70/3, C2/340/70/5, C2/600/70/3, C2/600/70/5, C2/1000/70/3, C2/1000/70/5, C2/1500/70/3, C2/1500/70/5, C2/2000/70/5
Temperature change rate, cooling (-70°C/ -50°C/ -40°C)13 valori
- - / 4 / 4.5 K/min
- C2/180/50/3 CO2, C2/600/50/4 CO2
- 3.8 / - / - K/min
- C2/180/70/3, C2/340/70/3
- 7.5 / - / - K/min
- C2/180/70/5
- - / 3.5 / 4 K/min
- C2/340/50/3 CO2
- 6.7 / - / - K/min
- C2/340/70/5
- 3.5 / - / - K/min
- C2/600/70/3
- 6 / - / - K/min
- C2/600/70/5, C2/1000/70/5
- 2.8 / - / - K/min
- C2/1000/70/3
- - / 10 / 12 K/min
- C2/1000/50/10 CO2
- 2.7 / - / - K/min
- C2/1500/70/3
- 5 / - / - K/min
- C2/1500/70/5
- - / 7 / 8.5 K/min
- C2/1500/50/7 CO2
- - / - / - K/min
- C2/2000/70/5
| Specificație | C2/180/50/3CO2 | C2/180/70/3 | C2/180/70/5 | C2/340/50/3CO2 | C2/340/70/3 | C2/340/70/5 | C2/600/50/4CO2 | C2/600/70/3 | C2/600/70/5 | C2/1000/70/3 | C2/1000/70/5 | C2/1000/50/10CO2 | C2/1500/70/3 | C2/1500/70/5 | C2/1500/50/7CO2 | C2/2000/70/5 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Testspace volume | 190 l | 335 l | 332 l | 600 l | 990 l | 1540 l | 2250 l | |||||||||
| External Dimensions (HxWxD) | 1830 × 900 × 1575 mm | 1830 × 900 × 1890 mm | 2040 × 1120 × 1925 mm | 2040 × 1420 × 2075 mm | 2040 × 1420 × 2600 mm | 2040 × 1420 × 3275 mm | ||||||||||
| Test Space Dimensions (H x W x D) | 750 × 580 × 450 mm | 750 × 580 × 765 mm | 950 × 800 × 800 mm | 950 × 1100 × 950 mm | 950 × 1100 × 1475 mm | 950 × 1100 × 2150 mm | ||||||||||
| Temperature range | -50 … 180 °C | -70 … 180 °C | -50 … 180 °C | -70 … 180 °C | -50 … 180 °C | -70 … 180 °C | -50 … 180 °C | -70 … 180 °C | -50 … 180 °C | -70 … 180 °C | ||||||
| Temperature change rate, cooling (-70°C/ -50°C/ -40°C) | - / 4 / 4.5 K/min | 3.8 / - / - K/min | 7.5 / - / - K/min | - / 3.5 / 4 K/min | 3.8 / - / - K/min | 6.7 / - / - K/min | - / 4 / 4.5 K/min | 3.5 / - / - K/min | 6 / - / - K/min | 2.8 / - / - K/min | 6 / - / - K/min | - / 10 / 12 K/min | 2.7 / - / - K/min | 5 / - / - K/min | - / 7 / 8.5 K/min | - / - / - K/min |
| Humidity range | 10 … 98 % R.H. | |||||||||||||||
Valoare comună modelelor peste care se întinde
- Fișă tehnicăPDF
- Absolute humidity controlPDF
- Saving energy with CO₂ refrigeration technologyPDF
- Climate Test Chambers ClimeEventPDF
- Fișă tehnică C2/180/50/3 (CO2)PDF
- Fișă tehnică C2/180/70/3PDF
- Fișă tehnică C2/180/70/5PDF
- Fișă tehnică C2/340/50/3 (CO2)PDF
- Fișă tehnică C2/340/70/3PDF
- Fișă tehnică C2/340/70/5PDF
- Fișă tehnică C2/600/50/4 (CO2)PDF
- Fișă tehnică C2/600/70/3PDF
- Fișă tehnică C2/600/70/5PDF
- Fișă tehnică C2/1000/70/3PDF
- Fișă tehnică C2/1000/70/5PDF
- Fișă tehnică C2/1000/50/10 (CO2)PDF
- Fișă tehnică C2/1500/70/3PDF
- Fișă tehnică C2/1500/70/5PDF
- Fișă tehnică C2/1500/50/7 (CO2)PDF