SO2 Humidity Chamber Kesternich

SO2 Humidity Chamber Kesternich

Kesternich testing simulates acid rain or industrial chemical exposure to evaluate the relative corrosion resistance of the coating, substrate, or part itself.

Kesternich testing simulates acid rain or industrial chemical exposure to evaluate the relative corrosion resistance of the coating, substrate, or part itself. Parts or panels are placed inside a specially designed chamber and are exposed to SO2 and condensation humidity before being evaluated for resistance to corrosion.

That makes the Kesternich Chamber unique.

  • Fully automatic operation: No manual intervention required during the tests
  • Multifunctionality: In addition to SO₂ tests, can also be used as a pure condensation humidity chamber
  • Precise control: Modern electronic gas dosing system and digital temperature control
  • Safety features: Hermetically sealed chamber, automatic air purging, pneumatic hood control
  • Space-saving options: Also available as a tabletop unit
  • User-friendliness: Intuitive controls, graphic temperature display, Ethernet connection

Fully Automatic, SO2 Gas Dosing Test Chamber.

A specific volume of SO2 gas, usually 0.2L, 1L or 2L is introduced into the test chamber, the chamber temperature is increased to a predetermined level and the relative humidity is maintained at condensation levels. After a set period, the chamber is vented and the temperature is allowed to decrease to ambient conditions.

Ascott’s Kesternich chamber does not require any human intervention whilst running the tests, its fully automatic design is controlled by a state of the art control system. The Kesternich chamber is designed to meet various gas dosing test standards as listed below and can also be used as a condensation humidity chamber.

Discover the advanced user interface of the Ascott Kesternich SO₂ Humidity Chamber in action.

The Ascott Kesternich SO₂ Humidity Chamber has an innovative touchscreen control system. The intuitive interface allows users to easily configure and manage all test parameters — including SO₂ gas dosing, humidity control, air purging, and drainage — all fully automated.

Designed for efficiency and repeatability, the system ensures consistent and accurate corrosion testing results.

Fully complies with the following standards.

ASTM D2247

NFT 30-077

ASTM G87

ISO 6270-2 CH

SFW 2,0,S

ISO 22479

ISO 3231

VDA 621-421

ISO 11503

NFT 30-055

BS3900/F9

EN ISO 6988

ISO EN 1096 B

DIN 50017 KK & KTW

DIN 50018

ISO EN 1096 C

Installation requirements.

Water: Deionized/distilled for topping up humidity generator.

Air: Clean dry & oil free, 4.0 to 6.0 bar (58-87psi) with 240 Ltrs (8.5cu.ft) per minute flow

Exhaust: 3m (10ft) exhaust pipe is required which should be terminated outside building (see accessory ACC58).

Drain: 3m (10ft) drain pipe is required which should be terminated into floor level drain (see accessory ACC58).

Operating environment conditions: Indoors, with ambient maintained at +18 to +23°C (+64 to 73°F), 85% max RH (non-condensing).

Options

  • Guardian Angel (ACC05): The Guardian Angel comprises a modular light alarm that can now be fitted to the top of all Ascott iP chambers. When the chamber is in daily use in the laboratory it allows the occupants of the area to see the status of the chamber visually at a distance depending on the colour of the status light: GREEN: Everything is operating correctly. AMBER: The chamber has a small problem which will eventually need attention but is still operational. RED: The chamber operation has been stopped and needs immediate attention.
  • Logging software (ACC121): This software allows logging of chamber variables and editing of chamber programs using a graphical interface. It also allows for remote programming, pause, start, stop. This software is suitable for logging a maximum of 3 chambers at any one time, providing all chambers are connected to the same network, either via an Ethernet hub or router.
  • Wall Mounted SO₂ Detector (ACC128): When working with SO₂ then safety is obviously paramount for all users of such equipment. We can offer a wall mounted detector that is powered by mains. The unit is to be fitted on or near to the chamber to help detect leaks proactively with the chamber or gas bottles.
  • Water Pressure booster (ACC61): An externally mounted water pump that can be used in order to boost the customer fed incoming water pressure.

Available models

  • KH300
  • KH300TT